Steam Sterilizer GK-100-3: specifications

Home cosiness

Sterilizer steam GK-100-3 TZMOI (Tyumena factory of medical equipment and instruments) is used for disinfection of medical linen, utensils, tools, laboratory devices. The increased power of the heating element contributes to the formation of a steam jacket, the performance of which is almost twice as high as that of the predecessor. The unit is equipped with a drying system for processed linen and materials for dressing. The main components of the equipment are a steam generator and a sterilization chamber.

steam sterilizer gk 100 3

Device

The steam sterilizer GK-100-3 includes the following elements:

  1. Sterilization compartment. The chamber is a cylindrical vessel, the front part of which is equipped with a lid. The middle part of the part has double walls, the space between which aggregates with the steam generator tube, forming a jacket for guaranteed heating of the chamber before the steam inlet, providing heating of the components in the drying mode.
  2. A gasket made of polymer material is placed in the groove of the chamber ring between it and the lid.
  3. The cover of the working compartment consists of a traverse, a round plate of a spherical type, a casing, a locking and screw mechanism.
  4. The locking device includes a handle and a bolt. Turn the handle to close or open the camera.
  5. Non-return valve - serves to pass air and steam exclusively in the direction of the condenser. It, in turn, is connected to a steam generator and equipped with a coil.
  6. Steam generator. The element is intended for the production of steam, which is used for sterilization.
  7. Connecting pipes and flanges.

Steam sterilizer ГК-100-3: technical characteristics

Below are the main technical indicators of the unit under consideration:

  • The maximum operating steam pressure is 0.22 MPa.
  • The supplied current is a variable three-phase parameter.
  • The frequency is 50-60 Hz.
  • The rated voltage is 380 V.
  • The required power is 16 kVA.
  • Sterilization modes - 2 ranges.
  • Length / width / height - 1,33 / 0,6 / 1,5 m.
  • Weight - 280 kg.
  • The average working life is more than 10 years.
  • Operating temperature (1/2 mode) - 132/120 degrees Celsius.

sterilizer steam gk 100 3 specifications

Operating principle

The steam sterilizer GK-100-3 is switched on vianetwork by means of a switch, after which the voltage is applied to the transformer, as indicated by the lamp indicator. After filling the steam generator with water, the voltage is transferred to the relay winding, which is activated, breaking the contact in the lamp circuit and the magnetic starter.

Inclusion of electric tubular heaterstype is carried out by means of the provided toggle switch. After it is set to "On", the process of generating steam in the generator is activated. When the temperature and pressure reach the maximum values, the relay switches off the magnetic starter and the heating elements. The stability of the steam parameters is maintained in the automatic mode, thanks to the magnetic starter, oriented to work in the specified operational parameters.

steam sterilizer gk 100 3 tsmoi

Safety while working with steam sterilizer GK-100-3

The manufacturer recommends that the following rules related to the safe operation of the instrument:

  • Persons who have received appropriate training are allowed to service the appliance.
  • Before starting work, the equipment must be earthed.
  • After 5-6 operating cycles, the valve with the stem should be raised to prevent it from boiling.
  • Periodically check the safety valve at the time of operation.
  • Observe the readings of the instrumentation (when the arrow enters the red zone - immediately turn off the equipment).

It is forbidden to maintain and operate the steam sterilizer GK-100-3:

  • Operate the equipment until the instruction manual is fully understood.
  • Switch on the device without grounding.
  • Leave the appliance unattended while working.
  • Perform sterilization in the event of a malfunction or failure of the instrument gauges.
  • Allow the formation of steam in the working chamber with a half-open bolt.
  • Open the latch of the chamber without leveling the pressure in it.
  • To repair the steam sterilizer GK-100-3, which is under voltage.

 sterilizer steam horizontal gk 100 3

Preparation for work

Before starting the operation of the equipment under consideration, it is recommended to perform a number of mandatory actions. Among them:

  • Inspection of the device after transport and unpacking for defects and damage.
  • Check of completing elements, cleaning of the device from special greasing, processing of a working chamber by hot water.
  • Installation of the TKMOI sterilizer GK-100-3 in a room equipped with sewerage, a water supply, a source of three-phase alternating current (380 V, 50 Hz).
  • Connecting the device, its grounding and testing.
  • The filter element must be filled not tightly, to avoid its clogging.

Exploitation

The procedure for working with the equipment under consideration is as follows:

  • Close the drain and open the fill valve.
  • Fill the steam generator with water to the level of the control mark.
  • The arrows are displayed on the manometer by adjusting with a screwdriver according to the pressure regimes.
  • The activation toggle switch is activated, and the warning light comes on.
  • Load processed clothing or other accessories into the sterilization compartment.
  • The control knob rotates clockwise to the mark with the desired operating parameters.
  • After the formation of the expected pressure, a safety valve opens, which contributes to a drop in the index.
  • During operation, the time from the beginning of operation to the end of the use of the device is noted (from 20 to 45 minutes, depending on the selected mode).
  • After the procedure is completed, open the valve to remove condensate.
  • Before each new cycle of work, check the presence of water in the steam generator at the proper level.
  • Switching off the equipment is carried out by deactivating the tumbler switch on and closing the drain cock after removing the water and condensate residues.

steam sterilizer gk 100 3 manufacturer

Service

Preventive and repair work of the horizontal steam sterilizer GK-100-3 is carried out by trained specialists. The list of works and manipulations includes such procedures:

  • Check the rubber seal for leaks.
  • Monitoring the ground of the steam generator.
  • Check the removal of residual water after work.
  • Cleaning of the heating elements and the water indicator glass every three months.
  • Regular check of sensors, pipelines, electrodes, covers for integrity and tightness.
  • Cleaning the spool and valves from residual products.

sterilizer steam gk 100 3 repair

Malfunctions and their elimination

Most often, the equipment in question can work with violations for the following reasons:

  1. The manometer does not display the realpressure in its presence in the working chamber. It is necessary to switch off the unit, check the siphon tube for contamination, and send the meter for inspection. The main cause of the malfunction is clogging of the siphon pipeline and damage to the manometer mechanism.
  2. Increase in pressure above the norm. In this case, you should replace the pressure gauge, since the main cause of the problem lies in its failure.
  3. No operation of the safetyvalve. It is necessary to carefully lift the rod of the device, remembering the possibility of getting a burn due to the high temperature in the working compartment. The reason is that the spool is attached to the seating seat.
  4. If the heating time exceeds 25-30 minutes, check the heating elements and replace them if necessary.
  5. </ ol </ p>